- It adopts an original imported 7.0-inch 800*480 pixel true-color TFT-LCD large-screen color display, with a human-sensing touchscreen keyboard, a fully Chinese operating interface, and an intuitive and user-friendly software interface. The interface is rich in content, easy to operate, and features Chinese character prompts, high detection sensitivity, simple operation, fast testing speed, and good repeatability.
- The instrument has test result storage, printing, and calculation functions. The sample determination process is automatically controlled by the instrument, with stirring and measurement completed automatically in approximately 60 seconds, directly displaying the results. The fully sealed titration bottle avoids contact between reagents and personnel, and also avoids the influence of environmental humidity. The instrument features a Chinese LCD display and automatic calculation and printing functions, capable of printing percentage content, sample number, tester, and experiment date. The instrument’s accuracy is: ① When the water content is less than 10 micrograms, the measurement error is less than 2 μg of water; ② When the water content is between 10 and 1000 micrograms, the measurement error is ≤2 μg of water; ③ When the water content is above 1000 micrograms, the measurement error is ≤0.2% (excluding injection error).
- The core host uses a TI AM3354 processor with a Cortex-A8 core and a 1GHz clock speed; the operating system is Windows Embedded Compact7 real-time industrial control system. This completely abandons the bare-bones microcontroller without a core or operating system, truly modernizing instrument control and ushering the instrument into a new era of intelligence.
Instrument Features:
- Capable of micro-analysis of low-content samples with high sensitivity.
- Can store 20,000 experimental results; employs unique alternating balance isolation detection technology, making balance detection faster, more accurate, and more stable; the electrolysis circuit uses advanced PWM pulse width modulation technology, significantly reducing overall power consumption.
- User-friendly human-machine interface with a touchscreen interface.
- Clock display (year, month, day, hour, minute, second), retains information even after power failure.
- Printer: Embedded thermal printer, 36 characters, Chinese character output; quieter, faster, and clearer printing.
- Program-controlled, allowing direct digital input from the interface to adjust the stirring speed.
- Multiple formula selections, automatically changes display units (ug, mg/L, ppm, %), and can be memorized.
Instrument parameters:
Display: 7.0-inch LCD color touchscreen
Measurement Range: 0.0001% (1ppm) to 100%
Sensitivity Threshold: 0.01ugH2O
Accuracy: ±2μg for 5μg-1mg, 0.2% for above 1mg (excluding injection error)
Test Results: Printed
Power: Less than 60W
Operating Environment: Temperature 15℃-28℃, Humidity <65%
Dimensions: 385×290×280 (mm)
Power Supply: AC220V±10% 50Hz±5%
Weight: Approx. 8kg
Meets the following criteria:
- GB/T7600 Determination of Moisture Content in Transformer Oil During Operation (Coulometric Method)
- GB6283 Determination of Moisture Content in Chemical Products – Karl Fischer Method (General Method)
- SH/T02346 Determination of Water Content in Light Petroleum Products (Coefficient Method)
- SH/T0255 Determination of Moisture Content in Additives and Additive-Containing Lubricating Oils (Coefficient Method)
- GB/T11133 Determination of Water Content in Liquid Petroleum Products (Kal Fischer Method)
- GB/T7380 Determination of Water Content in Surfactants (Kal Fischer Method)
- GB10670 Determination of Trace Moisture in Industrial Fluoromethanes – Karl Fischer Method
- GB/T606 General Method for Determination of Moisture in Chemical Reagents – Karl Fischer Method
- GB/T8350 Denatured Fuel Ethanol
- GB/T8351 Ethanol Gasoline for Vehicles
- GB/T 3776.1 Determination of Moisture Content in Pesticide Emulsifiers
- GB/T 6023 Determination of Trace Moisture in Industrial Butadiene – Karl Fischer Coulometric Method
- GB/T 3727 Determination of Trace Moisture in Industrial Ethylene and Propylene
- GB/T 7376 Determination of Trace Moisture in Industrial Fluoroalkanes
- GB/T 18619.1 Determination of Water Content in Natural Gas – Karl Fischer Coulometric Method
- GB/T 512 “Determination of Moisture Content in Lubricating Greases”
- GB/T1600 “Determination of Moisture Content in Pesticides”
- GB/T11146 “Determination of Water Content in Crude Oil (Karl Fischer Coulometric Method)”
- GB/T12717 “Test Methods for Industrial Acetates”
- GB/T5074 “Determination of Moisture Content in Coking Products”
- GB/T18826 “1,1,2-Tetrafluoroethane for Industrial Use (HFC-134a)”
- Meets the technical requirements for the determination of moisture content in pharmaceuticals using the Karl Fischer method as specified in the National Pharmacopoeia
- GB/T260 “Determination of Moisture Content in Petroleum Products”